The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HRG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.Sample1:21:41:81:16serum(n=5)88-105%94-102%90-103%86-103%EDTA plasma(n=5)85-97%84-97%86-96%88-98%heparin plasma(n=5)86-100%82-100%81-96%82-99%
manual-link-3
Matrices listed below were spiked with certain level of HRG and the recovery rates were calculated by comparing the measured value to the expected amount of HRG in samples.MatrixRecovery range(%)Average(%)serum(n=5)87-9993EDTA plasma(n=5)88-10295heparin plasma(n=5)85-9991